SPEC STD / SPEC STD Vis/NIR

​• Wavelength range: 380-1050 nm
​• Optical Resolution: 1.4 nm
​• 600 grooves/mm grating blazed at 750 nm
​• Order sorting filters included

Ranging from 380 to 1050 nm the SPEC STD Vis/NIR was designed with a 25 µm slit for good resolution and sensitivity, a 600 grooves/mm grating blazed at 750 nm for maximum efficiency on the NIR region and, as all STD spectrometers, includes a Sony 2048 pixels linear CCD array, order sorting filters, SMA connector for optical fibers and a Enhanced Sensitivity version as an option.

The SPEC STD Vis/NIR is suitable for Absorbance, Transmittance, Reflectance, and Irradiance in its detection range.

In the image below (left) shows the emission spectrum of a Mercury Argon Lamp acquired with this configuration. The detailed spectrum in the right shows the well resolved spectrum of the doublet 840.821 and 842.465nm.

 

Wavelength range 380-1050 nm
Slit 25 µm
Grating 600 lines @ 750 nm
Resolution 1.4 nm
Detector 2048 pixels CCD array Sony
Pixel width x Height 14 x 200 µm
Optical design Symmetrical Czerny-Turner
Order sorting filters Yes
Focal length 50 mm
Dark noise (RMS) 37
Dynamic range 1771
Signal/Noise 350
Linearity 99.99 %
Integration time 2 ms - 214 s
QE (%) @ peak 40%
Peak wavelength 450 nm
Pixel well depth 90000
PRNU 5%
A/D converter 16 bits
Interface Mini USB 2.0
Trigger In/Out
Trigger delay In
Fiber connector SMA 905
Weight 660 g
Dimensions 116 (L) x 93 (D) x 55 (H)
Collecting Lens Optional (ES Version)

 

The image below shows the spectra of a Light Source (LS-DW) from Sarspec, obtained with all four configurations from the SPEC STD family.

 

                     

Product Part number
SPEC STD Vis/NIR SP-STD-Vis-NIR
Enhanced sensitivity option SP-STD-Vis-NIR-ES

 

REQUEST A QUOTE